Laser sensor head - meter for power and energy of single pulses, for pulsed lasers: 40 mW - 20 W
- energy: 100 mJ - 20 J
- cooling: water
- wavelength range: 190 - 250 nm
- 3 years warranty
This laser sensor can be chosen in 3 connection options (only one per sensor):
- to an external laser power and/or energy meter (independent measuring console, but with PC connectivity via USB)
- direct to PC or controller via USB
- direct to PC or controller via RS-232
Ask also for OEM sensors for integration in your system or machine with lasers.
Full specifications LPT-20-UVC-D40 (-U / -RS)
Power Mode | |
Max. Average Power | 20 W |
Max. Intermittent Power (1) | 30 W |
Min. Power | 40 mW |
Power Resolution | 1 mW |
Noise Equivalent Power (NEP) | 1 mW |
Response Time (0-90%) | 4 sec |
Power Calibration Uncertainty | 3% |
Power Linearity (2) | 1% |
Single Shot Energy Mode | |
Max. Energy (with 100 ms pulse) | 20 J |
Min. Energy | 100 mJ |
Energy Resolution | 1 mJ |
Energy Calibration Uncertainty | 5% |
Absorber Specs | |
Aperture | 40 mm |
Type | UVC |
Absorber Spectral Range | 0.19 - 0.25 m |
Calibration Spectral Range | 0.19 - 0.25 m |
Max Power Density (3) | 40 W/cm |
Max Energy Density (3) | Single Pulse: (a) 10ms pulse width: 15 J/cm |
General Characteristics | |
Cooling | Convection |
Weight | 0.9 kg |
Dimension | 100 x 100 x 55 mm |
Cable lenght - connector | 2.5 m - USB |
Stand and Post | Light Duty Stand Included |
Notes | |
(1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (3). Damage thresholds also depend on power level. Please see damage graphs for more details. | (a). For repeated pulses, please see volume absorber damage graphs. |
Meet here all our laser power and energy sensors, splitted into their two main groups:
- standard laser power and energy sensors-meters
- BLINK: high speed laser sensors, for ultrafast lasers (picoseconds, femtoseconds).
If you use lasers you might also need:
- laser safety eyewear
- other laser protection products
- laser beam profile analyzers