Laser power sensor head - meter: 6 W - 3 KW
- cooling: water
- 3 years warranty
This laser sensor can be chosen in 3 connection options (only one per sensor):
- to an external laser power and/or energy meter (independent measuring console, but with PC connectivity via USB)
- direct to PC or controller via USB
- direct to PC or controller via RS-232
Ask also for OEM sensors for integration in your system or machine with lasers.
Full specifications LPT-W-3000-D55-HPB (-U / -RS)
Power Mode | |
Max. Average Power | 3 kW |
Max. Intermittent Power (1) | 4.5 kW |
Min. Power | 6 W |
Power Resolution | 1 W |
Noise Equivalent Power (NEP) | 0.25 W |
Response Time (0-90%) | 6 sec |
Power Calibration Uncertainty | 5% |
Power Linearity (2) | 2% |
Single Shot Energy Mode | |
Max. Energy (with 100 ms pulse) | n.a. |
Min. Energy | n.a. |
Energy Resolution | n.a. |
Energy Calibration Uncertainty | n.a. |
Absorber Specs | |
Aperture | 55 mm |
Type | HPB |
Absorber Spectral Range | 0.19 - 11 m |
Calibration Spectral Range | 0.19 - 2.1 m, 2.94m, 9 - 11 m |
Max Power Density (3) | 1.8 kW/cm @2kW |
Max Energy Density (3) | 5ms pulse width: 36 J/cm 10s pulse width: 1.2 J/cm 10ns pulse width: 0.3 J/cm |
General Characteristics | |
Cooling | Water (a) |
Weight | 2.3 kg |
Dimension | Ø 148 x 55 mm |
Cable lenght - connector | 5 m - USB / 5 m - RS232 |
Stand and Post | Heavy Duty Stand Included |
Notes | |
(1). 2 minutes max (2). Detector centrally irradiated @50% of useful surface. (3). Damage thresholds also depend on power level. Please see damage graphs for more details. | (a). Water 5 liter/min (@ 22C); admissible rate of temperature variation |
Meet here all our laser power and energy sensors, splitted into their two main groups:
- standard laser power and energy sensors-meters
- BLINK: high speed laser sensors, for ultrafast lasers (picoseconds, femtoseconds).
If you use lasers you might also need:
- laser safety eyewear
- other laser protection products
- laser beam profile analyzers